Figure 10. Amplifier Board Adjustment Locations
B. Adjust Fine DC Balance (R304)
1. Remove the shorting plug from the Logic Board.
2. Reinsert the shorting plug into the two holes near the end
of the Logic Board. (Figure 9.)
3. Reenter Cal Mode by pushing either of the two buttons that
protrude through holes in the Logic Board.
4.
5HPRYHWKH%1&FDEOHIURPWKHSUHFLVLRQ WHUPLQDWRU
connected to the DMM.
5. Short the output BNC connector by reconnecting the cable
to the SHORT connector on the AP033/AP034 Calibration
Fixture. The SHORT connector is the only BNC connector
on the AP033/AP034 Calibration Fixture that does not have
corresponding input pins for the probe tip. It is located on
the end of the board beyond the Common Mode
Terminated connector.
6. In Cal Mode, the EFFECTIVE GAIN indicators serve as an
adjustment indicator. The EFFECTIVE GAIN indicators are
located on the Logic Board (Figure 11). It may be
necessary to hold the boards in your hands to see the
indicators while making the adjustment.
R322C
Gain
R226
Coarse DC Bal
R330
Offset
R304
Fine DC Bal
Summary of Contents for AP034
Page 1: ... QVWUXFWLRQ 0DQXDO H UR 3 FWLYH LIIHUHQWLDO 3UREH REVISION D JANUARY 2000 ...
Page 6: ...YL ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 10: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 26: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 68: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 72: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH Figure 13 Typical CMRR Graphs ...
Page 76: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...