7. Using a low capacitance screwdriver, adjust Fine DC
Balance (R304) until the
÷
10 EFFECTIVE GAIN indicator
lights (Figure 11).
Figure 11.
÷
11 Effective Gain Indicator
8. Disconnect the BNC cable from the short connector.
Remove the shorting plug from the logic board.
C. Adjust Low Frequency CMRR (R7)
1. Leave the BNC cable connected to the probe end of the
ProBus extension cable. Connect a 50
%1& 7KURXJK
Terminator on the other end of the BNC cable. Connect the
male end of the 50
WHUPLQDWRUWRFKDQQHORQWKHVFRSH
The scope end of the ProBus extension cable should
remain connected to channel 1 of the scope, although there
is no signal cable connected.
2. Connect a BNC cable from the output of the Function
Generator to the Channel 2 input of the oscilloscope.
3. Connect a second BNC cable from the Function Generator
SYNC OUT to the external trigger input of the oscilloscope.
4. Set the Function Generator waveform to Sine and the
frequency to 70 Hz.
AP033 Logic Board
÷
10 Indicator:
Adjustment correct
Effective Gain Indicators
Adjustment too far
counter-clockwise
Adjustment too far
clockwise
Summary of Contents for AP034
Page 1: ... QVWUXFWLRQ 0DQXDO H UR 3 FWLYH LIIHUHQWLDO 3UREH REVISION D JANUARY 2000 ...
Page 6: ...YL ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 10: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 26: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 68: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 72: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH Figure 13 Typical CMRR Graphs ...
Page 76: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...