9.3 Warning
The RS 232/485 interface recognizes the following warning messages.
Table 29: RS 232/485 warning messages
Code
Meaning
501
Internal communication overloaded.
502
Unexpected reset. Contact LAUDA Service if the warning occurs multiple times.
503
Disconnection detected.
There was no communication for a period longer than the preset timeout.
Valid for the Variocool and PRO product lines.
508
Bus system faulty. Contact LAUDA Service if the warning occurs multiple times.
509
Unknown module connected.
510 – 532
Software for specified component outdated. Contact the LAUDA department.
V1
Interface module LRZ 926
37 / 43