ID Function
Unit
Command
155 Pressure limitation set point with active through-flow control
(MID flow controller must be connected and equipped with an inte-
grated pressure sensor)
[bar]
OUT_SP_10_X.X
159 Actual value of external pressure control variable (via interface)
[bar]
OUT_PV_06_XX.XX
Table 20: Cooling
ID Function
Unit
Command
23 Cooling mode: 0 = off / 1 = on / 2 = automatic
[–]
OUT_SP_02_XXX
Table 21: Safety
ID Function
Unit
Command
34 Timeout communication via interface (1 – 60 seconds; 0 = Off)
[s]
OUT_SP_08_XX
72 Activate Safe Mode
[–]
OUT_MODE_06_1
Table 22: Control parameters
ID Function
Unit
Command
38 Control parameter Xp
[–]
OUT_PAR_00_XX.X
40 Control parameter Tn (5 – 180 s; 181 = Off)
[s]
OUT_PAR_01_XXX
42 Control parameter Tv
[s]
OUT_PAR_02_XXX
44 Control parameter Td
[s]
OUT_PAR_03_XX.X
46 Control parameter KpE
[–]
OUT_PAR_04_XX.XX
48 Control parameter TnE (0 – 9000 s; 9001 = Off)
[s]
OUT_PAR_05_XXXX
50 Control parameter TvE (5 = Off)
[s]
OUT_PAR_06_XXXX
52 Control parameter TdE
[s]
OUT_PAR_07_XXXX.X
54 Correction limitation
[K]
OUT_PAR_09_XXX.X
56 Control parameter XpF
[–]
OUT_PAR_10_XX.X
60 Control parameter Prop_E
[K]
OUT_PAR_15_XXX
Table 23: Control
ID Function
Unit
Command
58 Set point offset
[K]
OUT_PAR_14_XXX.X
66 Control in control variable X: 0 = internal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2 (only for Integral)
[–]
OUT_MODE_01_X
V1