Integral XT
112
Starting up
YAWE0028 / 24/06/2016
Example:
Transfer of setpoint 30.5 °C to the thermostat with address 15.
Computer
Thermostat
“A015_OUT_SP_00_30.5“CR
ð
ï
“A015_OK“CR
7.17.5 Write commands (Data commands to the thermostat)
Command
Explanation
OUT_PV_05_XXX.XX
External temperature to be set through the interface.
OUT_SP_00_XXX.XX
Setpoint transfer with up to 3 places before the decimal point and up to 2 places
behind.
OUT_SP_01_XXX
Pump power level 1 to 8.
OUT_SP_02_XXX
Cooling operating mode cooling (0 = OFF / 1 = ON / 2 = AUTOMATIC).
OUT_SP_04_XXX.X
TiH outflow temperature high limit.
OUT_SP_05_XXX.X
TiL outflow temperature low limit.
OUT_SP_06_X.XX
Set pressure (with pressure control)
OUT_PAR_00_XXX
Setting of the control parameter Xp.
OUT_PAR_01_XXX
Setting of the control parameter Tn (5...180s; 181 = Off).
OUT_PAR_02_XXX
Setting of the control parameter Tv.
OUT_PAR_03_XX.X
Setting of the control parameter Td.
OUT_PAR_04_XX.XX
Setting of the control parameter KpE.
OUT_PAR_05_XXX
Setting of the control parameter TnE (0...979s; 980 = Off).
OUT_PAR_06_XXX
Setting of the control parameter TvE (0 = OFF).
OUT_PAR_07_XXXX.X
Setting of the control parameter TdE.
OUT_PAR_09_XXX.X
Setting of the correction limitation
OUT_PAR_10_XX.X
Setting of the control parameter XpF.
OUT_PAR_14_XXX.X
Setting of the setpoint offset.
OUT_PAR_15_XXX
Setting of the control parameter PropE
OUT_MODE_00_X
Master keyboard: 0 = free / 1 = locked (corresponds to “KEY”).
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external analogue / 3 = external serial.
OUT_MODE_03_X
Command remote control keyboard: 0 = free / 1 = locked
OUT_MODE_04_X
Setpoint offset source: 0=normal / 1=ext. Pt / 2=ext. analog / 3=ext. serial.
START
Switches the device on (after Standby).
See safety information (
STOP
Switches the device into Standby (pump, heater, cooling unit OFF).
RMP_SELECT_X
Selection of program (1...5) to which the further instructions apply. When the device
is switched on, program 5 is selected automatically.
RMP_START
Start the programmer.
RMP_PAUSE
Hold (pause) the programmer.
RMP_CONT
Restart the programmer after pause.
RMP_STOP
Terminate the program.
RMP_RESET
Delete the program (all Segments).
RMP_OUT_00_XXX.XX_XXXXX_XXX.
XX_X
Sets a programmer segment (temperature, time, tolerance and pump level). A seg-
ment is added and appropriate values are applied to it.
RMP_OUT_02_XXX
Number of program loops: 0 = endless / 1...250.