Integral XT
132
Maintenance
YAWE0028 / 24/06/2016
List of the fuses in the devices
From XT 490 W upwards:
Control fuse F3
à
T 0A2 (slow blow) Order no. EES 069.
Single-phase alternating cur-
rent units
Order no.
Circuit boards
UL 533 (mains)
UL 555-9 (power supply)
for all devices
---
F5/6/7
à
T10A0
(slow blow) EEF 026
XT 150
LWP 112; 230 V; 50 Hz
F1
à
T10A0
(slow blow) EEF 026
F3, F4
à
see table below
(
þ
XT 150
LWP 512; 200 V; 50/60 Hz
LWP 812; 208-220 V; 60 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 250 W
LWP 113; 230 V; 50 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 250 W;
LWP 513; 200 V; 50/60 Hz
LWP 813; 208-220 V; 60 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 350 W
LWP 117; 230 V; 50 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 350 W
LWP 517 ; ; 200 V; 50/60 Hz
LWP 817; 208-220 V; 60 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 350 HW
LWP 119; 230 V; 50 Hz
F1
à
T10A0
(slow blow) EEF 026
XT 350 HW
LWP 519; 200 V; 50/60 Hz
LWP 819; 208-220 V; 60 Hz
F1
à
T10A0
(slow blow) EEF 026
+
For the PCB UL 555 and the fuses F3 and F4, consider the following:
−
Only use UL fuses (listed according to UL 248-14!
Table fuses for single-phase alternating current units
View of PCB UL 555 (
þ
page 136)