Integral XT
104
Starting up
YAWE0028 / 24/06/2016
Command
dynamic heat limit
−
Open the device parameter menu with
the softkey Menu .
−
With the cursor keys continue to:
Change
à
Control
à
dynamic heat
limit
.
−
The window shown adjacent appears.
Select with the cursor key
Start
,
End
or
Set value
and confirm with
.
−
Enter the required limit temperature.
−
Accept the change with
,
−
or quit the window with
without mak-
ing any changes.
Example:
Start
= 250 °C
End
= 300 °C
Actuating signal
= 50 %
7.15.7.3
Dynamic control of heating power
With insufficient flow rates at the heaters there is a risk that the heat transfer liquid is locally overheated.
This may lead to earlier aging, oil cracking at silicone oils (depolymerisation) or bubbling up. At small
pump levels the heating power is regulated automatically by the dynamic heating power control. Gener-
ally up from pump level 5 the maximum heater capacity is available at all devices.
This function can not be adjusted or disabled.
Actuating
signal
in %
100
50
0
200
250
300
T in °C
maximum actuating
signal
Pump Menu
End
T
set
T
fix
dynamic heat limit
Start
250°C
End
300°C
Set value 50
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Start
250,
0
Min: 0°C
Max: 299°C