background image

 

DigiCal DCS 2

 

 

 

YAME0017/ 09.10.03 

Explanation of signs 

 

Explanation of signs: 

 

Danger: 

This symbol is used where there may be injury to 
personnel through incorrect handling. 

 

Note: 

Here spezial attention is drawn to some aspect. May 
include reference to danger. 

þ

 

Reference: 

Refers to additional information in other sections. 

Summary of Contents for DCS2

Page 1: ... P O Box 1251 Germany 97912 Lauda Königshofen valid from serie B02 09 03 YAME0017 Phone 49 0 9343 503 0 Fax 49 0 9343 503 222 e mail info lauda de Internet http www lauda de Operating instruction DigiCal Digital Thermometer DCS 2 ...

Page 2: ......

Page 3: ...be operated by technically qualified personnel Do not start up the equipment if it is damaged or leaking the supply cable is damaged Switch off the equipment and pull out the mains plug for servicing or repair before moving the equipment Have the equipment serviced or repaired by properly qualified personnel only The Operating Instructions include additional safety notes which are identified by a ...

Page 4: ...sical standard references mO1 13 2 4 3 3 DIN Ccalibration DIN IEC 60751 mEE 13 2 4 4 Printer port activation deactivation OpDl 14 2 4 5 Selection Channel activation Bgl 15 2 4 6 Analogue output c B 16 2 5 Recalling the memory data HOLD MAX MIN AVE 17 2 6 Measuring rate FAST Modus 17 2 7 AUTO OFF Function 17 3 Power supply 18 4 Error codes 18 5 Technical data 19 6 RS232 interface protocol 20 7 Conn...

Page 5: ...planation of signs Danger This symbol is used where there may be injury to personnel through incorrect handling Note Here spezial attention is drawn to some aspect May include reference to danger þ Reference Refers to additional information in other sections ...

Page 6: ...unctional elements YAME0017 09 10 03 1 Control and functional elements DCS 2 front view Measuring port for channel 2 Temperature Measuring probe channel 1 firmly connected Control panel Keys ON OFF switch Righthand adjustable base ...

Page 7: ... 03 Handling Introduction 7 DCS 2 rear view Analogue outlput 0 1V for channel 1 red plus Analogue output 0 1V for channel 2 red plus RS 232 Interface 9 pins Power supply input Temperature measuring probe channel 1 firmly connected ...

Page 8: ...ase connect first the measuring device with the suitable mains plug unit see picture on the back side of the device and then connect the mains plug with the mains supply socket þ3 Power supply The measuring probe 1 is firmly connected to CH 1 If you would like to connect a second measuring probe Plug the second probe into the socket CH 2 of the measuring device The connection sockets on the housin...

Page 9: ...nts such as like the temperature unit calibration of probes deactivation of channels and others are resulting from the structure of the menu You will reach into the main menu by pressing ENTER MENUE With the up and down keys you can choose your required menu item Push ESC to be back in the measuring mode up and down keys key ESC key ENTER MENUE Menu structure Press the key accordingly to pass thro...

Page 10: ...g unit and press ENTER MENUE to confirm Press ESC to get back to the measuring mode Display Keys Function Tlhr ENTER MENUE 1 Tlhr 2Tlhr up and down keys select the channel 1 or 2 1 Tlhr ENTER MENUE 1 Tlhr B 1 Tlhr E up and down keys select the measuring unit B or E ENTER MENUE To keep the amended value ESC To escape without retention 2 4 2 Temperature Difference Khl1 To display the difference betw...

Page 11: ...t the electronics This probe data can be read as a 2x4 digits code on the handle of the LAUDA DigiCal probes and results from the 2 point comparison method mO0 mO1 Calibration by physical standard references with the best possible system accuracy for a user s calibration calibration with comparative standards 1 point 2 point or 3 point comparison possible mEE Standard characteristic curve with a m...

Page 12: ...tepping to the next number use the down key If the requested number is complete then press ENTER MENUE to confirm Now on the bottom of the display appears a small 2 and after this number a second four digit number is displayed For changing the number please follow the manual as before Press ESC to get back to the measuring mode Important After confirming mO0 by pressing ENTER MENUE the function mO...

Page 13: ...s to select the requested sign Rh _ reference standard shows a negative measuring value below 0 00C Rh _ reference standard shows a positive measuring value above 0 00C Press ENTER MENUE to confirm On the second line of the display appears Ec Use the up and down keys to select the requested range Ec below 100 00 C Ec 0 above 100 00 C Press ENTER MENUE to confirm Now the display shows Please enter ...

Page 14: ... Use the up and down keys to select OpDl and press ENTER MENUE Use the up and down keys to activate ml or deactivate mEE the requested adjustment and press ENTER MENUE to confirm Press ESC to get back to the measuring mode Display Keys Function OpDl ENTER MENUE OpDl ml mEE up and down keys activation or deactivation ENTER MENUE To keep the amended value ESC To escape without retention ...

Page 15: ...hows ml for activation or mEE for deactivation Use the up and down keys again to activate ml or deactivate mEE the requested measuring channel and press ENTER MENUE to confirm Press ESC to get back to the measuring mode In case that one channel would not be connected to a probe the main display would indicate mODl in the relevant line There must be at least one active channel Display Keys Function...

Page 16: ...gits after decimal point cO two decimal digits cO one decimal digit decimal point moves one digit to the right Press ENTER MENUE to confirm On the second line of the display appears Rh 0 Use the up and down keys to select the requested range Rh reference standard shows a negative measuring value below 0 00C Rh 0 reference standard shows a positive measuring value above 0 00C Press ENTER MENUE to c...

Page 17: ...ied on Clearing the memory MAX MIN AVE Press CLEAR key once to erase the stored maximum minimum and average from memory On the display appears B p After erasing the memory the instrument automatically switches back to measuring mode indicating the actual measured value again 2 6 Measuring rate FAST Mode Press FAST key one to change the measuring rate Now the measuring rate is app 4 measurements pe...

Page 18: ...s required 230V 50 60Hz Ref No EBE 042 90 260V 50 60Hz Ref No EBE 043 4 Error codes By displaying the following error codes the instrument support the operation of the instrument Dppmp Description mODl no probe or wrong probe is connected rmGh too high exceeding of the measuring range rmKm too low underflow of the measuring range ...

Page 19: ...5 C 200 C 0 1 Resolution 200 bis 200 C with 0 01 C resolution 200 bis 400 C with 0 1 C resolution Outputs RS232 Interface serial Printer can be connected 2 analogue outputs 0 1 V 11 bit scalable Connectors DIN 8 pole Allowable operating temp 0 C 40 C Display 2 line LCD Housing plastic ABS Dimensions 200 x 80 x 200 mm BxHxL Weight 950 g Power supply Power Supply 230V 50 60Hz or optional 90 260V 50 ...

Page 20: ...e data the instrument lock automatically the keyboard To release the keyboard you have to send the command 0 hex The datas would be sent in the following format Datentyp String he length of the string depends on the instrument a two channel or one channel instrument measuring value 1 unit measuring value 2 unit LF CR 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 Example Turb...

Page 21: ...DigiCal DCS 2 YAME0017 09 10 03 Connector layout 21 7 Connector layout 7 1 Connector layout Pt100 4 wire Measuring channel 1 2 7 2 RS232 PC adapter cable Connector layout Measuring channel 1 2 ...

Page 22: ...be is labelled obviously with the relevant code If you change the probe you have to enter the code into the measuring instrument We recommend to write down both values in order to avoid any problems especially if the probe is positioned at a place difficult to access or if there are several external probes in use Ref no Appellation Serial number Calibration data Ref no Appellation Serial number Ca...

Page 23: ...d 300 C LMZ 053 PC adapter cable RS232 with a sub D plug for the DCS 2 and a 9 pin sub D coupler for the PC EKS 037 Windows PC Data Recording Software WinTherm Plus on CD ROM LSDM2002 Plug In Mains Supply for 90 260VAC 50 60Hz including plug adaptors for Euro plug USA UK Japan EBE 043 10 Ordering spares To avoid questions delay and delivery of incorrect items please advise always your instruments ...

Page 24: ......

Page 25: ... LAUDA voir plaque signalétique Typ Type Type Serien Nr Serial no No de série mit folgendem Medium betrieben wurde was used with the below mentioned media a été utilisé avec le liquide suivant Darüber hinaus bestätigen wir daß das oben aufgeführte Gerät sorgfältig gereinigt wurde die Anschlüsse verschlossen sind und sich weder giftige aggressive radioaktive noch andere gefährliche Medien in dem Ge...

Reviews: