DigiCal DCS 2
YAME0017/ 09.10.03
Technical data
19
5. Technical
data
Type:
DigiCal DCS 2
Probe Ports 1 / 2
Pt100
Measuring range
-200...+450°C
Accuracy
Pt 100 temperature probe with
manufacturer calibration
-200...–100°C: 0.05°C
-100...+150°C: 0.03°C
+150...+200°C: 0.05°C
>200°C: 0.1%
Resolution
-200 bis +200°C with 0.01°C resolution
+200 bis +400°C with 0.1°C resolution
Outputs
RS232 Interface (serial Printer can be connected)
2 analogue outputs 0-1 V (11-bit scalable)
Connectors
DIN 8-pole
Allowable operating temp.
0°C ... +40°C
Display
2-line LCD
Housing
plastic (ABS)
Dimensions
200 x 80 x 200 mm (BxHxL)
Weight
950 g
Power supply
Power Supply 230V; 50/60Hz or optional 90-260V; 50/60Hz
DCS 2 Ref. No.
LMD 018 (230V)
LMD 919 (90-260V)