DigiCal DCS 2
12
Handling / Introduction
YAME0017/09.10.03
2.4.3.1
Encoded Calibration -
mO0
☞
This calibrating function is only needed in case you would add another probe to the
second measuring channel.
Use the up and down keys [
] to select
mO0
. Press [ENTER/MENUE] to confirm the requested
setting. On the bottom of the display appears a small 1, followed by a four-digit number
(Hex-Code /0..F). The number can be modified with the up key [
]. For stepping to the next number
use the down key [
]. If the requested number is complete then press [ENTER/MENUE] to confirm.
Now on the bottom of the display appears a small 2 and after this number a second four-digit number
is displayed. For changing the number please follow the manual as before.
Press [ESC] to get back to the measuring mode.
Important!
After confirming
mO0
by pressing [ENTER/MENUE] the function
mO0
(encoded calibration) is
activated, even though you leave the menu by pressing the [ESC] key
Display-indication with active calibration code
mO0
:
The
B@K
– segment and the small 1 on the left indicates to the user
that
mO0
is activated.