updates to the current loop.
3.3.
Advanced noise Is always set to 0.
3.4.
What Running Average do you need?
3.4.1.
Available options are 0(off) and from 2 seconds to 30 seconds in whole
second intervals.
4.
Error Reporting. ($CE and $NE)
Error reporting can be turned on or off ($NE). An enhanced feature is the addition of
consecutive error reporting. During turbulent surface condition momentary single return losses
can occur. By setting this, you can account for this condition and only report error that are true
in nature.
Note: For more information see serial commands $CE and $NE.
4.1.
Do you error reporting?
4.1.1.
Yes
$NE,0 (On)
4.1.2.
No
$NE,1 (Off)
4.2.
What kind of liquid surface is most common?
4.2.1.
No Turbulence $CE,3
4.2.2.
Slight Turbulence
$CE,10
4.2.3.
Moderate Turbulence $CE,100
4.2.4.
Very Turbulence
$CD,255
Note:
Any whole number between 1 to 255 can work but the above values work in
most cases.
5.
Sensing Environment ($DM)
The TruSense sensor can identify multiple targets, if found. In stilling wells, light steam or fog
the last target reporting will mitigate the possibility of early returns from these obstacles.
Note: For more information see serial command $DM
5.1.
What measurement conditions?
5.1.1.
Open vessel $DN,5 (First Target)
5.1.2.
Stilling well $DM,7 (Last Target)