Section 6 – Serial Interface
Page 47
Remote Trigger (RUN) Request
$GO
,
n
<CR><LF>
$GO
The request identifier.
n
Number of iteration.
If number of iterations is omitted, it executes same as previous run.
<CR>
A carriage return.
<LF>
An optional line feed character.
Instrument Response:
$OK
*csum
<CR><LF>
$OK Message
identifier.
*csum
CRC16 checksum of data string up to the asterisk.
Examples:
$GO,0 Continuous measurements
$GO,3 3
measurements
Remote Trigger (STOP) Request
$ST<CR><LF>
$ST
The request identifier.
<CR>
A carriage return.
<LF>
An optional line feed character.
Instrument Response:
$OK
*csum
<CR><LF>
$OK Message
identifier.
*csum
CRC16 checksum of data string up to the asterisk.
Examples:
Request:
$ST<CR><LF>
Response: $OK*0744