2.2 Product Deployment Requirements
IR Fever Warning System is a kind of precise temperature measurement and imaging camera which has certain
requirements for deploying application. The installation and deployment diagram is shown in the figure 1-3. Specific
notes are as follows:
2.2.1 Installation Environment
➢
The device should be placed indoors and the ambient temperature should be kept relatively
constant
(
It’s necessary to setup a tent or shelter, such as good sealing environment and keep the
environment stable if the actual application scene is outdoors
)
;
➢
The position of the device head needs to be well-lit, and it should be avoid in the position of backlight
;
➢
There should be no air-conditioning outlets within 3 meters of the location of the equipment head;
➢
The field of view monitored by the device head must not be disturbed by strong external heat
sources such as light and sun
;
2.2.2 Head Layout
➢
Reasonably arrange the position of the IR236 head, and the downward plan angle is 13°
-
15°
;
➢
Reasonably arrange the detection position and range of personnel. The
position of IR236 camera from the forehead of the person under test is
3.0-10.0 meters, and the best test position is 5.0 meters
;
2.2.3 Tested person requirements:
➢
After outdoor sun exposure or after strenuous exercise, test in an indoor environment for 5 minutes
after peaceful adjustment
;
➢
People enter indoors from outdoor scenes. It is recommended to walk through the corridor buffer
for at least 10 seconds before stepping into the detection position
;
➢
Make sure that the entire face of the person under test is in the field of vision of the device, and that
the face and forehead cannot be blocked (hats, scarves, sunglasses, masks, antipyretic stickers
and other objects that may block the face. It is recommended to remove them in advance and adapt for
0.5-1 minutes carry out testing
;