SR1 Operation
267
© 2014 Stanford Research Systems
2.8.8
IMD Panel
The
IMD
Quick Measurment panel sets up SR1's analyzers, generators, and displays to perform any
one of the three standard intermodulation distortion (IMD) measurements: SMPTE, CCIF, and DIM.
Cabling Configuration
The output of SR1's analog or digital generator is connected to the EUT's input. The output of the EUT is
connected to SR1's analog or digital inputs. The number of inputs and outputs and the connectors used
are specified in the
Setup Panel
.
Generator Configuration
The output waveform is set to the
IMD
waveform of the specified type. For SMPTE measurements the
"high frequency" is set to the value entered in the panel while the low frequency defaults to 60 Hz. For
CCIF measurements the center frequency is set to the specified value while the difference frequency
defaults to 1 kHz. For DIM measurements the tone frequencies are fixed by the standard. The default
tone frequencies set up by the quick measurement can always be manually overridden on the
appropriate panel.
Analyzer Configuration
One or both analyzers (depending on the number of output channels specified in the
Setup Panel
) are
configured as
IMD Analyzers
. The measured distortion products are selectable from the Analysis section
of the panel.
Free Run Configuration
A barchart display is created for each output channel showing the instantaneous IMD distortion of the
EUT.
Sweep Configuration
A
graph
will be created for each channel showing IMD vs.amplitude for the selected range of amplitudes.
Log spacing can be selected for both the frequency and amplitude sweeps. The settling parameters of
the THD+N measurement can be adjusted directly from the "Settling" tab of the panel.
Summary of Contents for SR1
Page 5: ...Part I Getting Started Audio...
Page 7: ...Getting Started 7 2014 Stanford Research Systems...
Page 12: ...SR1 Operation Manual 12 2014 Stanford Research Systems...
Page 27: ...Part II SR1 Operation Audio...
Page 258: ...SR1 Operation Manual 258 2014 Stanford Research Systems...
Page 272: ...SR1 Operation Manual 272 2014 Stanford Research Systems on the amplitude sweep...
Page 289: ...SR1 Operation 289 2014 Stanford Research Systems...
Page 290: ...Part III SR1 Reference Audio...