VitalsBridge Directions for Use
33
rev. A.5
Enter the value shown on the patient monitor and SUBMIT. Verify that this reading is from the real vital signs monitor
and not a reading from a patient simulator.
The process of entering in a SpO
2
value is repeated until enough points are found for an accurate calibration reading.
Once enough calibration points are measured, the calibration data is automatically saved and selected.
Unused calibrations may be removed by highlighting the desired calibration to remove and selecting REMOVE.