15
5
Fully insert canister holder into the LSU.
6
Insert the opposite side of the Serres Vacuum
Connector into the canister.
7
Turn the operating knob to 500+ mmHg.
8
Close the Patient inlet with your finger and
simultaneously push the middle of the lid
down. The lid is properly installed when
vacuum has reached 500 mmHg.
9
Release the patient inlet and make sure that
the bag is fully inflated.
10
Turn off the LSU.
11
Fasten the Suction Tubing on the Suction
Tubing Holder
12
Place the LSU on charge (see section 6.4)
5.4 Serres Suction Bag System
Quick Reference
Correct assembly of the tubing
is illustrated on the right side of
the LSU.
5.5 After Use
1
Inspect all parts of the LSU for damage and
excessive wear.
Replace parts if necessary.
The Serres Vacuum Connector should be
replaced regularly.
2
Place an unopened Serres Suction Tube on
the side of the LSU.
3
Perform Device Test according to section 7.
4
Place the LSU on charge see section 6.4.