4 LSU Reusable Canister System
4.1 Overview
Aerosol Filter
Vacuum inlet
User Interface
Canister Holder
PATIENT Inlet
Patient Suction Tubing
Suction Catheter
Adapter
Holder for Water
Bottle/Patient
Suction Tubing.
Water Bottle
Canister Lid
with Float Ball
Battery Door
Release Arm
Canister
4.2 Parts
Gasket
Vacuum seal
PATIENT Inlet
Canister
Aerosol Filter
Angled Connector
Patient Suction Tubing
Canister Lid w/Float Ball
Suction Catheter Adapter
English
7
The LSU Reusable Canister System consists of:
• A 1000 ml transparent plastic Canister.
• A Canister Lid w/Float Ball
• An Aerosol Filter with tubing
The Canister Lid contains a Float Ball that shuts
off the vacuum if the canister is full or the LSU
tips over. To restore the vacuum, remove the
Angled
Connector from the VACUUM Inlet. The Float
Ball will be released and the Angled Connector
can be reattached,
Note:
The LSU Reusable Canister System must
not be used without the Aerosol Filter or the
Float Ball.
Decontamination of reusable parts must be
performed in accordance with section 4.6.