![laerdal LSU Directions For Use Manual Download Page 14](http://html2.mh-extra.com/html/laerdal/lsu/lsu_directions-for-use-manual_3279708014.webp)
English
14
1
Plug Patient Inlet with plug provided on the
lid. Disconnect the Serres Vacuum Connector
from the canister.
2
Press the release arm while sliding the holder
towards you to loosen or completely remove
canister.
3
Remove suction bag from canister
4
Safely dispose of the Serres Suction bag
according to local protocols.
5
Clean the LSU if needed according to section
6.2.
6
Clean the Serres canister if needed according
manufacturers instructions for use.
Note:
Clean the Serres Vacuum Connector by
wiping with a damp cloth or a sponge.
Do not autoclave or attempt to disassemble.
The Serres Vacuum Connector should be
replaced regularly.
5.3.4 Assembly of the Serres Suction Bag system
1
Place the canister in a canister holder in an
upright position.
2
Partially insert the canister holder into the
LSU.
3
Unfold the Suction Bag and place inside the
canister. Suction bag has to be installed into
equal size canister
4
Insert the Serres Vacuum Connector into the
LSU Vacuum Inlet.