Guidance and manufacturer’s declaration – electromagnetic immunity
The Laerdal Suction Unit (LSU) is intended for use in the electromagnetic environment specified below. The customer
or the user of the LSU should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic
tile. If floors are covered with
synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst.
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
± 1 kV differential mode
± 2kV common mode
± 1 kV differential mode
± 2kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
< 5% UT
(> 95% dip in UT) for 0,5
cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
< 5% UT
(> 95% dip in UT)
for 5 sec.
< 5% UT
(> 95% dip in UT) for 0,5
cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
< 5% UT
(> 95% dip in UT)
for 5 sec.
Mains power quality should be
that of a typical
commercial or hospital
environment.
Power Frequency
(50Hz/60Hz) magnetic
field. IEC 61000-4-8
3A/m
3A/m
NOTE: UT is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Laerdal Suction Unit (LSU) is intended for use in the electromagnetic environment specified below. The customer
or the user of the LSU should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The LSU uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The LSU is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies