18 

 
17. 

Trouble Shootings 

 

Error Indicator 

Possible cause 

Corrections 

 
 
 
Power on , 
No display 

Power supply is not 
connected 
 

Check the power supply system to see if there is 
voltage on the line 

Plug    has no access to 
socket 
 

Plug in    firmly 

The power switch has not 
been turned on. 
 

Turn on the power switch on the right   
side of the unit 

The fuse is broken 
 

Replace fuse with new one of same specification 

  Circuit occurs mall function   
of power box   

Notify distributor for repair service 

 
Actual temperature 
is higher than the 
set point,    high 
temperature alarm 
is activated 
 
 
 
 

Unit has not yet reached the 
required (constant) 
temperature. 
 

Wait a moment and observe 

Temperature setting is at 
the blind area of 
temperature control   

Open the ventilation hole 

Improper setting of 
refrigeration parameter 

Set the refrigeration parameter 

to be “0.5”and close 

the ventilation opening. 

The ventilation fan is broken  Notify the distributor to replace the fan 
Malfunction occurs with 
refrigeration system 

Notify the distributor to repair the refrigeration 
system 

 
 
Actual temperature 
is lower than the 
set point. 
low temperature   
alarm is activated 
 
 

Unit has not yet reached the 
required ( constant)   
temperature 

Wait a moment and observe 

The circulation of cold air is 
excessive 

Close(a part of) the ventilation opening 

The ventilation fan is broken  Notify the distributor to replace the fan 
The heater does not work 

Notify the distributor to repair the heater 

 
Actual temperature 
is fluctuating and 
will not be stable 

Improper setting of 
refrigeration parameter 

Refer to Users Guide and reset the refrigeration 
parameter 

The door (lid) is not closed 
firmly 

Close the door (lid) firmly 

Malfunction occurs with the 
control circuit 

Notify distributor for repair service 

Temperature is 
constant out of 
control 

Malfunction occurs with the 
control circuit 

Notify distributor for repair service 

 
The oscillation of 
platform is unstable 
 

The platform is in imbalance 
due to a spoiled object 

Remove the object , clear and clean the chamber 

The equipment is not placed 
horizontally 

Adjust the left-back foot    leveler install      the 
equipment    in a proper way. 

Malfunction occurs with 
control circuit 

Notify distributor for repair /service 

 
The shaking 
platform does not 
work 
 
 

The door switch has not yet 
made contact 

Check the door to see if it is closed firmly 

The platform is blocked with 
an object at the bottom   

Remove the platform and clear the object and clean 
the inside 

The belt is broken 

Notify the distributor to replace the belt 

Malfunction occurs with 
control circuit 

Notify distributor for repair service 

The oscillation of 

Malfunction occurs with 

Notify distributor for repair service 

Summary of Contents for ZWY-111B

Page 1: ...User s Manual This Manual applies to Modes ZWY 111B ZWY 211B ZWY 111G ZWY 211G ZWY 111C ZWY 211C ZWY 111D ZWY 211D ZWF 111 and ZWF 211 Economic Horizontal Shaking Incubator...

Page 2: ...agriculture It is widely used for cell culturing hybridization cell aeration and solubility studies Etc Since the date of your purchase of this product after sale service will always be close to you t...

Page 3: ...r is needed 6 A surge protector is recommended to avoid power related faults 7 In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power...

Page 4: ...orm is still moving do not open the lid too much or too long as this might affect the temperature inside 8 The incubator must be kept away from volatile flammable explosive liquids or gases 9 Please k...

Page 5: ...__ 10 9 Temperature Speed Alarm Settings _____________________________ 11 10 Refrigeration Defrost Settings ________________________________ 12 11 Temperature Calibration______________________________...

Page 6: ...r in refrigerated units only has an independent overload protection CFC free refrigerant which causes no damage to ozone layer Applies on ZWY 2 ZWF 2 models Large doublefold tempered glass window and...

Page 7: ...64 82 82 66 100 ml 64 82 82 66 250 ml 40 45 52 45 500 ml 28 33 38 32 750 ml 23 24 26 23 1000 ml 15 15 22 15 2000 ml 11 12 11 3000 ml 8 11 11 5000 ml 6 6 6 Test tube 18ml 208 208 208 208 Inner dimensio...

Page 8: ...n Hold this button for 2 seconds to turn the unit on off the standby mode Temperature button Press this button to display the actual temperature value and the set temperature in turns Speed button Pre...

Page 9: ...amber is deviated over the preset alarm limit the indicator flashes and the alarm beeps Heating indicator When the heater is connected with power supply this indicator will be light This indicator wil...

Page 10: ...eep unscrewing screwing the adjusting bolt on the front left corner with the spanner until the unit is elevated leveled and sits firmly on the four blocks 4 2 The shaking mechanism of all reciprocal s...

Page 11: ...the current status of temperature retaining Keep Temp Mode function press Increase or Decrease to change the setting 6 2 Press Set confirm to exit to the home page 1 Display current control mode and t...

Page 12: ...imer page press Increase or Decrease to change the value if necessary If Timer is set as 0 the unit will keep running continuously 7 2 Press Set confirm to exit to the home page Special Attention 1 Th...

Page 13: ...ess Increase or Decrease to change the time press Set confirm to save and switch to year setting page press Increase or Decrease to set the year and press Set confirm again to save and switch to the m...

Page 14: ...Increase or Decrease to change the value 9 3 Press Set confirm to exit to the home page Special Attention 1 Deviation alarms can only be activated when unit is running and once actual temperature or...

Page 15: ...ss the Set Confirm button to confirm and store the changed value 10 5 Explanation for setting auto defrosting cycle a Defrost Timer is the time interval between two cycles For example if this is set a...

Page 16: ...it will start manual defrost for set defrosting period To abort press button and hold for 3 seconds Minimum time interval between two manual defrost cycles is 1 min Attention 1 The defrosting cycle wi...

Page 17: ...ture should have changed due to the calibration Waiting for another one hour to let the temperature stabilize again at 37 C and check the thermometer value again and calculate the new difference f if...

Page 18: ...tton if the symbol A is displayed on the lower left corner of the screen it means the non volatile memory function has been active 13 Setting of communication address for RS485 connection Opt If the R...

Page 19: ...on within 3 seconds the remaining operating time will be cleared to zero Press the Start Stop button once more and the unit starts again to count down from the preset operation time 15 5 While the uni...

Page 20: ...through the opening 16 5 Use the Allen key to loosen the A central fixing nut 16 6 Use the 8 10 spanner to hold the B adjusting bolt as still in the meantime loosen the C safety nut with 17 19 spanne...

Page 21: ...cold air is excessive Close a part of the ventilation opening The ventilation fan is broken Notify the distributor to replace the fan The heater does not work Notify the distributor to repair the heat...

Page 22: ...he equipment if it does not work Notify the distributor The equipment causes a strange loud noise The equipment is not placed horizontally Adjust the left back foot to make the equipment stable The fi...

Page 23: ...20 18 Electronic Control System...

Page 24: ...e the parts free of charge Your distributor will provide professional workmanship and will repair the unit without any charge When this one year warranty is expired we will continue to provide you wit...

Reviews: