9
7.
Temperature, Speed, Timer Settings
7.1. Press "Set/confirm", press the "Increase" button to "3", then press the "Set/confirm" again. The
screen shows the current temperature preset value, press "Increase" or "Decrease" to change the
value; press "Set/confirm" to save and switch to speed page, press "Increase" or "Decrease" to
change the value, and press "Set/confirm" again to save and switch to the timer page, press
"Increase" or "Decrease" to change the value if necessary.
If Timer is set as "0", the unit will
keep running continuously.
7.2. Press "Set/confirm" to exit to the home page
Special Attention!
1. This unit is equipped with "door switch". The temperature control
(main heater) will be temporarily disabled when lid is opened.
Therefore, always close the lid in time to avoid exceeded heat loss.
2. The timer can be set within 0-9999mins, the unit stops running
when timer is due. Acoustic alarm can be muted by pressing the
"Time" button.
3
Set
Temperature
Exit
Set Shaking
Speed
Set Timer