11
9.
Temperature & Speed Alarm Settings
9.1. Temperature and speed alarm is activated when temperature or speed is deviated over limit. The
deviation alarm limit can be preset as below,
9.2. Press "Set/confirm", press the "Increase" button to "5", then press the "Set/confirm" again. The
screen shows the current "temperature deviation" value, press "Increase" or "Decrease" to change
the value; press "Set/confirm" to save and switch to "speed deviation" setting page, press
"Increase" or "Decrease" to change the value.
9.3. Press "Set/confirm" to exit to the home page
Special Attention!
1. Deviation alarms can only be activated when unit is running, and
once actual temperature or speed has reached its preset value and
stabilized for a certain period of time.
2. Deviation alarm is activated when the actual parameter value
exceeds preset value
±
preset deviation alarm value. The "Alarm"
indicator is lighted, but can be muted by pressing "Run/Stop" button.
3 Deviation alarm can be automatically deactivated once if the
parameter value is back within the deviation tolerance range.
5
Set
Temperature
Deviation
Exit
Set Speed
Deviation