Refrigeration & Defrost Settings 

 

(Only for refrigerated models) 

This  unit  is  equipped  with  automatic  and  manual  defrosting  functions.  With  proper  settings  of 
refrigeration function and auto-defrosting cycle, this unit can be  used for continuous operation at low 
temperature settings. Please note defrosting function will only be activated when actual temperature at 
10

  or below. 

 

 

 

 

 

 

 

10.1 Press the "Set/Confirm" 

button, then press the Increase button to “6”, next press the "Set/Confirm" 

button to display the preset refrigeration function on the screen, SV flashes. Press the "Increase" 
or "Decrease" button to switch the refrigeration function "On/Off" . 

10.2 Press "Set/Confirm" to confirm setting and enter the page of defrost timer setting, press "Increase" 

or "Decrease" to set the time interval between two defrosting cycles. Set range from 1-240 hours. 

10.3 Press "Set/Confirm" to confirm setting and enter the page of defrost period setting. Press 

"Increase" or "Decrease" to set the time length of each defrost cycle. Set range from 1-24 minutes. 

10.4 When setting is finished, press the Set/Confirm button to confirm and store the changed value.   

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

10.5 Explanation for setting auto defrosting cycle: 

a)  "Defrost Timer" is the time interval between two cycles. For example, if this is set as "6", the 

unit will defrost once every 6 hours.   
 
To find out the suitable "defrost timer" setting, the frosting time duration should be counted from 
start of the low-temperature operation to the point that the actual temperature start bouncing up 
due to the frosting inside the evaporation chamber. Then set the "defrost timer" as somewhere 
between 1/10-1/5 of the frosting time duration. For example, if the frosting time duration is 20 
hours,  the  "defrost  timer"  can  be  set  from  2-4  hours.  High  relative  humidity  will  require  more 
frequent defrosting cycle. 
 

b)  "Defrost Period" is the time duration of each defrosting cycle. For example, if this is set as "3", 

the unit will run a defrosting cycle for 3 minutes.   
 
When the ambient temperature is higher than 25

, like in summer, the recommended setting 

is 30 seconds, otherwise, 1 min is considered as adequate. As ambient conditions may vary, 
therefore,  always  try  to  find  the  most  appropriate  settings  for  your  own  particular  ambient 
condition. 

Set 

Refrigeration 

Function       

(Cool 

Parameter) 

Exit 

Set Defrost 

Timer 

Set Defrost 

Period 

Summary of Contents for ZWY-111B

Page 1: ...User s Manual This Manual applies to Modes ZWY 111B ZWY 211B ZWY 111G ZWY 211G ZWY 111C ZWY 211C ZWY 111D ZWY 211D ZWF 111 and ZWF 211 Economic Horizontal Shaking Incubator...

Page 2: ...agriculture It is widely used for cell culturing hybridization cell aeration and solubility studies Etc Since the date of your purchase of this product after sale service will always be close to you t...

Page 3: ...r is needed 6 A surge protector is recommended to avoid power related faults 7 In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power...

Page 4: ...orm is still moving do not open the lid too much or too long as this might affect the temperature inside 8 The incubator must be kept away from volatile flammable explosive liquids or gases 9 Please k...

Page 5: ...__ 10 9 Temperature Speed Alarm Settings _____________________________ 11 10 Refrigeration Defrost Settings ________________________________ 12 11 Temperature Calibration______________________________...

Page 6: ...r in refrigerated units only has an independent overload protection CFC free refrigerant which causes no damage to ozone layer Applies on ZWY 2 ZWF 2 models Large doublefold tempered glass window and...

Page 7: ...64 82 82 66 100 ml 64 82 82 66 250 ml 40 45 52 45 500 ml 28 33 38 32 750 ml 23 24 26 23 1000 ml 15 15 22 15 2000 ml 11 12 11 3000 ml 8 11 11 5000 ml 6 6 6 Test tube 18ml 208 208 208 208 Inner dimensio...

Page 8: ...n Hold this button for 2 seconds to turn the unit on off the standby mode Temperature button Press this button to display the actual temperature value and the set temperature in turns Speed button Pre...

Page 9: ...amber is deviated over the preset alarm limit the indicator flashes and the alarm beeps Heating indicator When the heater is connected with power supply this indicator will be light This indicator wil...

Page 10: ...eep unscrewing screwing the adjusting bolt on the front left corner with the spanner until the unit is elevated leveled and sits firmly on the four blocks 4 2 The shaking mechanism of all reciprocal s...

Page 11: ...the current status of temperature retaining Keep Temp Mode function press Increase or Decrease to change the setting 6 2 Press Set confirm to exit to the home page 1 Display current control mode and t...

Page 12: ...imer page press Increase or Decrease to change the value if necessary If Timer is set as 0 the unit will keep running continuously 7 2 Press Set confirm to exit to the home page Special Attention 1 Th...

Page 13: ...ess Increase or Decrease to change the time press Set confirm to save and switch to year setting page press Increase or Decrease to set the year and press Set confirm again to save and switch to the m...

Page 14: ...Increase or Decrease to change the value 9 3 Press Set confirm to exit to the home page Special Attention 1 Deviation alarms can only be activated when unit is running and once actual temperature or...

Page 15: ...ss the Set Confirm button to confirm and store the changed value 10 5 Explanation for setting auto defrosting cycle a Defrost Timer is the time interval between two cycles For example if this is set a...

Page 16: ...it will start manual defrost for set defrosting period To abort press button and hold for 3 seconds Minimum time interval between two manual defrost cycles is 1 min Attention 1 The defrosting cycle wi...

Page 17: ...ture should have changed due to the calibration Waiting for another one hour to let the temperature stabilize again at 37 C and check the thermometer value again and calculate the new difference f if...

Page 18: ...tton if the symbol A is displayed on the lower left corner of the screen it means the non volatile memory function has been active 13 Setting of communication address for RS485 connection Opt If the R...

Page 19: ...on within 3 seconds the remaining operating time will be cleared to zero Press the Start Stop button once more and the unit starts again to count down from the preset operation time 15 5 While the uni...

Page 20: ...through the opening 16 5 Use the Allen key to loosen the A central fixing nut 16 6 Use the 8 10 spanner to hold the B adjusting bolt as still in the meantime loosen the C safety nut with 17 19 spanne...

Page 21: ...cold air is excessive Close a part of the ventilation opening The ventilation fan is broken Notify the distributor to replace the fan The heater does not work Notify the distributor to repair the heat...

Page 22: ...he equipment if it does not work Notify the distributor The equipment causes a strange loud noise The equipment is not placed horizontally Adjust the left back foot to make the equipment stable The fi...

Page 23: ...20 18 Electronic Control System...

Page 24: ...e the parts free of charge Your distributor will provide professional workmanship and will repair the unit without any charge When this one year warranty is expired we will continue to provide you wit...

Reviews: