Chapter 4: Using Your Stoppering Tray Dryer
Product Service 1-800-522-7658
16
Original instructions
Power Switch
– Turns all AC power to the Tray Dryer ON or OFF.
POWER / STANDBY LED Indicator
– Indicates if the power switch is ON or OFF.
IN PROCESS LED Indicator
– This light will be ON if the Stoppering Tray Dryer is
running in either MANUAL or PROGRAM mode.
Stoppering Control Valve
– Controls the operation of the stoppering bladder. The white
dot on the valve knob indicates the position of the valve. Turning the valve to the
RAISE positon will inflate the bladder to perform the stoppering operation. Turning
the valve to the LOWER position will cause the bladder to deflate.
Vacuum Release Valve
- Vents the chamber to allow the chamber door to be opened.
Backfill Port
– A regulated gas cylinder may be connected to this port to allow the
introduction of a desired gas into the chamber after freeze drying. The port accepts
1/8” ID tubing.
Ethylene Oxide is not recommended for use in this Tray Dryer for
decontamination because of its hazardous and corrosive properties
.
Sample Probe Ports
– (3) sample probes may be plugged into the Stoppering Tray Dryer
to monitor the sample temperature during the freeze drying process.
Stoppering Bladder
– Can be inflated to perform vial stoppering
Operation Checklist
The following checklist should be followed prior to each use of your Stoppering
Tray Dryer:
1.
Using a soft, lint-free cloth, wipe the Stoppering Tray Dryer chamber to
remove any moisture or debris. Wipe door gasket to remove any debris or
other contaminants that could cause a vacuum leak. Vacuum grease is NOT
required on the door gasket to obtain a proper vacuum seal.
2.
Remove the collector chamber lid from the Freeze Dryer and ensure that the
collector chamber and drain line are free of water. Place the drain hose in a
suitable container to collect the condensate from the collector chamber and
insert the quick connect drain fitting into the quick connect drain coupling
located on the left hand side of the Freeze Dryer.
Freeze Dryer (Collector or
Vacuum) will not start if moisture is detected in the drain line.
3.
After completely draining the system, disconnect the quick connect drain
fitting from the quick connect drain coupling.
!