Chapter 4: Using Your Stoppering Tray Dryer
Product Service 1-800-522-7658
26
Original instructions
9.
Ampules may be flame sealed while connected to a valve by using a
sealing torch. Care must be taken not to burn the valve. An insulation
material placed between the valve and the torch is recommended.
10.
Shut off the Freeze Dryer, defrost the ice and drain the collector and dry.
See the Freeze Dryer User’s Manual.
Shelf Spacing Adjustment
Prior to loading the shelves with the product to be freeze dried, adjust the shelves
to the 3-shelf, 2-shelf or 1-shelf position. It is best to observe the shelf supporting
mechanism and practice while reading these procedures. In the 3-shelf position,
the spacing between the shelves will accommodate a serum bottle with the stopper
raised up to a maximum height of 75 mm (2.9 inches).
Minimum serum bottle height with lowered stopper for bottles on 3 shelves is 38
mm (1.5 inches).
For a 2-shelf system, grasp the top shelf and raise it until it latches in place. Then
raise the middle shelf approximately 1 inch until it latches in a position to allow a
two equally spaced shelf system. The spacing between shelves will accommodate
a bottle with the stopper raised up to a maximum height of 107 mm (4.2 inches).
Minimum serum bottle height with lowered stopper for a 2-shelf system is 56 mm
(2.2 inches).
For a single shelf system, raise both the top and middle shelves until they latch in
place in their top positions. The spacing for a single shelf system will
accommodate a bottle with raised stopper up to a maximum height of 196 mm
(7.7 inches), and the minimum height of a bottle with lowered stopper for
successful stoppering is 114 mm (4.5 inches).
Note location of bevel
on knob in vent position
Valve in Vent Position