Appendix C: Freeze Dryer Specifications
Product Service 816-333-8811 or 1-800-522-7658
62
Original instructions
Environmental Conditions
Indoor use only.
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II (Overvoltage
Categories per IEC 1010). Temporary voltage spikes on the AC input line
that may be as high as 1500V for 115V models and 2500V for 230V
models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.