Chapter 5: Using Your Freeze Dryer
Product Service 816-333-8811 or 1-800-522-7658
40
Original instructions
1.
Go to the Programs screen. You can access the Programs screen in two
different ways, both lead to the same location.
a.
Directly access the Programs screen by selecting the Programs icon
from the main menu bar at the top of the display.
b.
Go to the Drying Accessory screen, then select the PROGRAM
button at the bottom of the screen.
2.
Select a program from the saved programs list, or create a new program.
3.
With the desired program selected from the list, press the START button.
Shut Down
At the end of a run or when a sufficient amount of condensate accumulates on the
collector coil to obstruct the flow of vapor to the collector chamber, the Freeze
Dryer should be defrosted. Go to Home screen, press VACUUM, then STOP to
turn the vacuum pump OFF. Then release system vacuum by turning the plastic
knob on a valve (that has no sample attached) to the open position or opening the
vent valve on Tray Dryer or Clear Chamber. Press COLLECTOR, then STOP to
turn the refrigeration system OFF.
Defrosting
The following procedure should be followed when defrosting the collector coil:
1.
The hot gas defrost function can be used to speed up the defrosting process.
Defrost can be activated as follows: Go to Home screen, press COLLECTOR
button, press Defrost Options, then press START.
2.
To turn off Defrost: Go to Home screen, press DEFROST button, press STOP.
The defrost function will turn off automatically if the collector temperature
r60°C or if defrost has been running for 2 hours.
3.
If rapid defrost is desired, pour warm water over the collector coil.
Do not
allow the liquid to enter the vacuum port on the upper rear wall of the chamber.
4.
Place the drain hose in a suitable container to collect the condensate from the
collector chamber. Insert the quick connect drain fitting into the quick connect
drain coupling located on the left hand side of the unit.
5.
Flush the collector chamber with water and wipe chamber dry.
6.
Disconnect the quick connect drain fitting from the quick connect drain
coupling.
DISCONNECT THE QUICK CONNECT DRAIN FITTING
BEFORE STARTING THE VACUUM PUMP. FAILURE TO
REMOVE THE FITTING WILL RESULT IN A LARGE
VACUUM LEAK.
Utilization of acid requires immediate cleaning and
neutralization after defrost or physical damage to the collector
chamber and collector coil will result.
Do not attempt to chip ice from the collector coil as damage
may occur to the coil.
!