Chapter 4: Lyo-Works Operating System
Product Service 816-333-8811 or 1-800-522-7658
25
Original instructions
3.
Units –
Select your desired unit of measure for Temperature (°C , °F) and
Vacuum (mbar, Pa, Torr)
4.
Security
– A four digit security code can be set that will require users to enter
the code to gain access to selected screens. A security code will always be
required to enter the “Security” screen.
THE FACTORY DEFAULT SECURITY CODE IS “
1234
”
System
– If the System security feature is turned ON, users will be
required to enter the security code to gain access to the touch screen
controls when the screen is reactivated from a screen timeout or
power failure (screen timeout described in the Settings/Display
section).
Programs, Email Alerts, Calibration, Network
– When the
security feature for these screens is selected, users will be required
to enter the security code to gain access to each individual screen.
5.
Alerts –
If a network valid connection has been established (via the Settings /
Network screen) and Email Settings have been configured, alerts can be
emailed to selected users, by entering the appropriate email address(es) and
selecting the type of alerts that will be sent to each email address.