36
©2016. All Rights Reserved. Jeio Tech Co., Ltd
User can check the parameter setting value 9 using (Config button), and also change the parameter
setting value (by Up/Down button)
Lower limit: -10 ~ upper limit (by Celsius)
Upper limit: lower limit to 120 degrees (Celsius criteria)
The screen will be returned to the main screen, If the product is standby condition for 5 sec.
If the temperature limiter shown value is different with actual value (measured calibration temperature
device),
Please corrected in the following way;
a. Shut if the power of the product.
b. Press the Config button with Up/Down button, and turn on the main power.
c. About 4 sec later, every setting will be initialization and show OFFSET setup screen.
d. User can set OFFSET value by Config button.
e. Set the OFFSET value (-10~10
Celsius criteria
) by using Up/Down button
f. Save the Config value.
g. Supply the main power again to product, the product will be applied by changed OFFSET value.
4.8.2 Current Protection
Fuse : it can automatically shut off the current when overcurrent condition immediately.
4.8.3 Door Switch
The unit heater and fan will be stopping when door is open automatically.
Summary of Contents for IL3-15
Page 1: ...OPERATION MAUNAL IL3 series Model IL3 15 IL3 25 Manual No 220412L002 Version 0 0 ...
Page 6: ......
Page 7: ...1 Manual NO 220412L002 Version 0 0 1 0 Safety ...
Page 17: ...11 Manual NO 220412L002 Version 0 0 2 0 Functional Description ...
Page 19: ...13 Manual NO 220412L002 Version 0 0 2 3 Structure IL3 15 IL3 25 ...
Page 21: ...15 Manual NO 220412L002 Version 0 0 3 0 Installation ...
Page 26: ...20 2016 All Rights Reserved Jeio Tech Co Ltd 4 0 Operation ...
Page 43: ...37 Manual NO 220412L002 Version 0 0 5 0 Maintenance ...
Page 46: ...40 2016 All Rights Reserved Jeio Tech Co Ltd 6 0 Troubleshooting ...
Page 49: ...43 Manual NO 220412L002 Version 0 0 7 0 Jeiotech Software ...
Page 52: ...46 2016 All Rights Reserved Jeio Tech Co Ltd 8 0 Appendix ...