Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM
Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1)
5.2 AED Mode Configuration and Operation in wired mode
AED mode is available for acquiring images without any connection to the X-ray generator with an X-ray SYNC
cable
The way of image acquiring is as follows;
1)
Selection of AED mode
2)
Shoot the X-ray using hand switch of X-ray generator.
Make sure to follow operating environmental requirements (Temp: 10℃ ~ +40℃)
If you use AED Mode out of operating environmental requirements, unwanted
image can be acquired without X-ray image acquiring.
Do not hit or drop the equipment. Unwanted images may be acquired in the AED
Mode if it receives strong jolt.
If you use a Grid under general imaging condition (Dose) or image a thick object in
the AED Mode, the efficiency of X-ray transformation may be reduced about 0% ~ 2%
compared to the Manual Mode according to the thickness of the target.
If you image a thick object in the AED Mode with low X-ray tube voltage, an image may
not be acquired or horizontal line noise may occur.