When using the device, take the following precautions. Otherwise, problems may occur and the
device may not function correctly.
Image Backup
To avoid missing images which might result in the patient being exposed to additional doses of ra-
diation, it is important to backup images by filming or using a CD or DVD option. This should be done
as a routine operation for every patient.
If the hard disk of your workstation is about to be full, the operator should backup images and delete them
to make storage (memory) space on the hard disk for new information (patient).
User Limitations
The PEDRA-DMS (Detector Managing Software) has the technician mode and this mode should be
operated by the personnel who are qualified by the manufacturer
Disposal
Disposal of this product in an unlawful manner may have negative effects on human health and the
environment. When disposing of this product, therefore, be absolutely sure to follow the procedure which
is in conformity with the laws and regulations applicable in your area.
Before exposure
Be sure to check the device daily and confirm that it works properly.
Sudden heating of the room in cold areas will cause condensation to form on the equipment. In this case,
wait until the condensation evaporates before performing an exposure. If the equipment is used while
condensation is formed in it, problems may occur in the quality of captured images.
When an air-conditioner is used, be sure to raise/lower the temperature gradually so that a difference
between the temperature in the room and in the equipment does not occur, which prevent con-
densation.
During exposure
This device is not protected (sealed) against liquids such as blood and medicine in the operating room.
If necessary, wrap the equipment in a disposable cover when using.
Do not use the detector near devices generating a strong magnetic field. Doing so may produce image
noise or artifacts.
Cleaning and Disinfection