“More than just Digital”
Page
75
of
81
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM
Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1)
Electromagnetic Immunity
The PEDRA is intended for use in the electromagnetic environment specified below. The customer
or user of this device should assure that it is used in such an environment.
Phenomenon
Basic EMC
standard or test
method
Operating
mode
Port tested
Test Voltage
Test lev-
el/requirement
Mains terminal
disturbance
voltage
CISPR11:2015
Operating
AC Mains
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
Group1, Class A
Radiated dis-
turbance
CISPR11:2015
Operating
Enclosure
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
Group1, Class A
Harmonic Cur-
rent Emission
EN 61000-3-2:2014
IEC 61000-3-2:2014
Operating
AC Mains
230 V,50 Hz
Class A
Voltage change,
Voltage fluctua-
tions and
Flicker Emission
EN 61000-3-3:2013
IEC 61000-3-3:2013
Operating
Stand by
AC Mains
230 V,50 Hz
Pst: 1
Plt: 0.65
Tmax:0.5
dmax: 4%
dc: 3.3%
Electrostatic
Discharge Im-
munity
EN 61000-4-2:2009
IEC 61000-4-2:2008
Operating
Stand by
Enclosure
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
± 8 kV/Contact
± 2, ± 4, ± 8,
± 15 kV/Air
Radiated RF
Electromagnetic
Field Immunity
EN
61000-4-3:2006+A2:
2010
IEC 61000-4-3:2006
+A2:2010
Operating
Stand by
Enclosure
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
3 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Immunity to
Proximity Fields
from RF wireless
Communications
Equipment
EN
61000-4-3:2006+A2:
2010
IEC 61000-4-3:2006
+A2:2010
Operating
Stand by
Enclosure
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
Table 9 in
IEC 60601-1-2:
2014
Electrical Fast
Transient/Burst
Immunity
EN 61000-4-4:2012
IEC 61000-4-4:2012
Operating
Stand by
AC Mains
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
± 2 kV, 100 kHz
repetition frequency
SIP/SOP
± 1 kV, 100 kHz
repetition frequency
Surge Immunity
EN 61000-4-5:2014
IEC 61000-4-5:2014
Operating
Stand by
AC Mains
120 V, 60 Hz
220 V, 60 Hz
230 V,50 Hz
Line to Line
± 0.5 kV, ± 1 kV
Line to Ground
± 0.5 kV, ± 1 kV, ± 2
kV
Immunity to
Conducted Dis-
EN 61000-4-6:2014
IEC 61000-4-6:2013
Operating
AC Mains
100 V, 50 Hz
100 V, 60 Hz
3 V
0.15-80 MHz