Patient lift/Patient stander inspection log
Dat
e of
inspection
Code and
serial
number
of the
bod
y
support
unit
Information
regarding the
conditions of
the bod
y
support unit
Intervention
ty
pe
(indicate the
number of
the
paragraph
or note
dow
n the
intervention)
Faults
identified
Accessories
used
Date of next
inspection
Corrective
actions or
feedback
Identification
and Signature
of the
Inspection
Manager
69
11.13
Routine maintenance
To ensure that the device maintains its level of efficiency and characteristics, routine
servicing must be carried out on an annual basis at the premises of the authorised Support
Centres. To find out where the Support Centres are located, please contact your local
representative/dealer or
KSP Italia Srl
Via Dell’Artigianato, 1
Post code 06031 Bevagna (Perugia -Italy)
Tel. +39 0742 361947 - Fax +39 0742 361946
[email protected] - www.kspitalia.com
Routine maintenance includes the following:
- checking the electrical system
- checking the resistance with load cycle
- checking the drive buttons
- checking the actuator
- checking the wheel-movement system
- checking the state of wear of the wheels
- checking the safety devices and the status of the battery-charging unit
- checking battery efficiency levels
- checking the condition of the frame, lift boom and spreader bar
- checking the lifting system
- checking the mechanical transmission system
- checking the body support unit
- checking the fixing systems
- checking current dispersion
- checking braking efficiency
- checking of the load-bearing structure and lifting mechanism with relative connections,