26
6.2
Verifying the correct installation
Proceed as follows to verify that installation has been carried out correctly:
1.
Check that all the screws and bolts are tightened in the manner and to the torques indicated
in the paragraph relating to assembly.
2.
Check the operation of the mechanism for changing the width of the base by intervening on
the manual lever/pedal.
3.
Check the system used to raise/lower the electric/oil‐hydraulic patient lift by completing 5
movement cycles with no person on the lift, from the lowest point to the highest point,
checking the correct operation and that there is no friction or abnormal noises.
4.
If the previous check has a positive outcome, repeat the same test with a weight on the lift
equal to the maximum load, checking the correct operation and also that there is no friction
or abnormal noises.
5.
Check the operation of the caster wheels, ensuring that there is no clearance and that they
move freely.
6.
Cover 50 linear metres then make turn the lifter on itself 5 times in one direction and 5 times
in the opposite direction with no load. Repeat the operations with the maximum load,
checking for correct operation and also that there is no friction or abnormal noise.
7.
Place the maximum load on the device, brake the wheels and check that the device does not
move.
8.
Check the status of the body support unit used and the batteries.
9.
Check for the presence of the battery charger and its connection.
10.
Check the labelling and, if the fiscal document accompanying the device indicates the serial
number, check that this number matches the number on the label.
11.
Check that the instruction manual is included and verify the presence and condition of the
labels, as indicated in the relevant paragraph.
12.
Carry out all the checks and verifications listed in the maintenance section.
13.
Check that in the areas of use there are suitable access points and manoeuvring spaces.
14.
Record the installation checks in the register indicated in the maintenance paragraph with
the wording ‘VERIFICATION OF CORRECT INSTALLATION’.
27
7.
Technical specifications
Illustrated below are different device models with their constituent parts.
7.1
Models and components Item N305‐150
Weight of parts that can be disassembled for transport: base,
column, boom with spreader bar
1) Lifting actuator (150 kg) with emergency descent device
2) Handle
3) Push‐button control panel
4) Dewert control unit with emergency descent device
5) Foot pedal to open/close the arms
6) Structural support base
7) Support column
8) Lift boom with spreader bar
9) Removable battery
1
2
3
4
5
6
7
8
9
108
109