46
DO NOT USE THE DEVICE IF THE BATTERIES ARE NOT CHARGED TO AT LEAST 40% OF THEIR
FULL CHARGE. IF THE BATTERY IS DEPLETED, REPLACE IT WITH A CHARGED BATTERY OR
ENSURE THAT THE BATTERY IS CHARGED BEFORE STARTING ANY OPERATION.
Checklist prior to use:
a)
Condition of the mechanical parts;
b)
Condition of the sling;
c)
For electrical devices, check the battery charge, check that the battery is not depleted, and:
i.
for the DEWERT drive, check that the charge LED is not red and no acoustic
signal is emitted while the actuator is being activated;
ii.
for the LINAK drive, check that no acoustic alarm activates when the controls
are operated.
d)
For oil‐hydraulic devices, check there are no fluid leaks from the actuator;
e)
For electric devices, check the operation of the red lock ring located at the top of the
actuator for moving the device manually in an emergency;
f)
For electric devices, check that the emergency buttons operate properly;
g)
Check that the device operates correctly when moving the boom from the highest point to
the lowest point and vice versa;
h)
Check that the adjustment mechanism of the base works correctly;
i)
Check the condition and free movement of the caster wheels;
j)
Check the operation of the wheel brakes;
k)
Check that the bolts indicated as ‘coupling’ bolts in the diagram relating to bolt tightening
are in contact with the screw‐in surface and that there is no axial clearance;
l)
Check that the bolts indicated as tightened to torque or not indicated as ‘coupling’ bolts
are screwed in fully;
m)
Check that the locking pins of the patient standers are present and engaged correctly;
Checklist after use:
a)
Condition of the mechanical parts;
b)
Condition of the sling;
47
c)
For electric devices, check the battery status, check there has been no abnormal energy
consumption;
d)
For electric devices, check there are no smells relating to overheated electric material;
e)
For electric devices, check that the battery has not overheated;
f)
For hydraulic devices, check there are no fluid leaks from the actuator;
g)
Check that the bolts indicated as ‘coupling’ bolts in the diagram relating to bolt tightening
are in contact with the screw‐in surface and that there is no axial clearance;
h)
Check that the bolts indicated as tightened to torque or not indicated as ‘coupling’ bolts
are screwed in fully;
i)
Check that any bolt locking pins are present and engaged correctly.
While using the device, the safety and well‐being of the User is of the utmost priority.
Therefore, follow the safety instructions provided in this manual and the indications
contained in the description of this device.
The person being transferred on the device must be calm and relaxed at all times.
Make sure that the User makes no abrupt or jerky movements.
The User must not be holding any objects. Ensure that the garments or limbs of the User
do not hinder the lift movements.
The raising and lowering operations are to be carried out taking into account the heightened
sensitivity of the User.
With the User on the lift, the total weight of the lift itself increases.
During transport the Operator must always grip the handle of the device pusher with
both hands.
N.B.: all the commands are ‘hold‐to‐run’.
8.2
Use on flat surfaces
To move the device, push or pull it manually using the caster wheels.
128
129
Summary of Contents for N305-150
Page 41: ...Note Note 158 159...
Page 42: ......