4
PROFIBUS SETTINGS
34
VFC 200
www.krohne.com
06/2019 - 4005059401 - AD VFC200 PB-PA R01 en
C3.3.6.2 Time Constant
Set time constant for this function block.
C3.3.6.3 Error Behavior
Defines the behaviour of this function block in case of errors.
Setting:
Replace Value: The "replacement value" will be used as OUT value;
Hold Value: Last valid OUT value stored will be used as OUT value;
Ignore Error: OUT has the wrong calculated value and status "Bad" as
calculated"
C3.3.6.4 Replacement Value
Defines the value that replaces the measured value at this function block in
case of an error.
C3.3.7 FB Analog Input 5
C3.3.7.1 Measurement
Setting:
Volume Flow; Norm. Volume Flow; Mass Flow; Gross Power; Net Power;
FAD; Density; Temperature 1; Temperature 2; Pressure; Vortex Frequency;
Velocity; Specific Enthalpy; Specific Heat Capacity; Reynolds Number
C3.3.7.2 Time Constant
Set time constant for this function block.
C3.3.7.3 Error Behavior
Defines the behaviour of this function block in case of errors.
Setting:
Replace Value: The "replacement value" will be used as OUT value;
Hold Value: Last valid OUT value stored will be used as OUT value;
Ignore Error: OUT has the wrong calculated value and status "Bad" as
calculated"
C3.3.7.4 Replacement Value
Defines the value that replaces the measured value at this function block in
case of an error.
C3.3.8 FB Analog Output
C3.3.8.1 Measurement
Setting:
Temperature External; Pressure External
C3.3.8.2 Unit
Setting:
°
C;
°
F;
°
Rank; K; atm; bar; mbar; Pa; Pa (abs); kPa; Mpa; psi; psi (abs); torr;
g/cm
²
; kg/cm
²
C3.3.8.3 Error Behavior
Defines the behaviour of this function block in case of errors.
Setting:
Replace Value: The "replacement value" will be used as OUT value;
Hold Value: Last valid OUT value stored will be used as OUT value;
Ignore Error: OUT has the wrong calculated value and status "Bad" as
calculated"
C3.3.8.4 Replacement Value
Defines the value that replaces the measured value at this function block in
case of an error.
C3.3.9 FB Totalizer 1
(same menu structure as stated within C4.3)
C3.3.10 FB Totalizer 2
(same menu structure as stated within C4.4)
C4 Totalizer
C4 Totalizer
C4 Totalizer
C4 Totalizer
C4.3 FB Totalizer 1
C4.3.1 Set
Starts / stops the totalizer and defines its value if stopped.
Setting:
Totalize: Totalizing is started / is in progress (Normal operation of the
totalizer).
Reset: Totalizing is stopped, 0.0 is assigned to the totalizer.
Preset: Totalizing is stopped, the preset value is assigned to the totalizer.
Function
Setting / Description