COMMISSIONING / OPERATION
3
31
VFC 200
www.krohne.com
06/2019 - 4005059401 - AD VFC200 PB-PA R01 en
3.3.4 Filtering of "Single events"
As already mentioned before each "Event group" can consist of up to 32 "Single events". In some
special cases a "Single event" should not be taken into consideration. Using "filtering" this
"Single event" can be disabled
–
that means: if this "Single event" will occur the diagnosis event
bit of the "Event group" associated with this "Single event" will not be set.
Filtering of "Single events" using the VFC 200 PROFIBUS DTM:
Enter Menu "Parameters
→
Input
→
Diagnostics":
•
"Open" "Event group" required by "single click" on "+" in front of the name of the "Event
group".
•
All "Single Events" belonging to this group will be displayed.
•
Unmark all check boxes of "Single events" to be filtered.
•
Transfer new filter setting(s) to the PROFIBUS device.
Filtering of "Single events" using the VFC 200 PROFIBUS EDD:
Enter menu:
→
Online menu
Online menu
Online menu
Online menu: "Device
→
Setup
→
Device
→
Diagnosis mapping and filter
→
Diagnosis filter":
•
Within the "Diagnosis filter" area all "Single Events" will be displayed which might be filtered.
•
Unmark all check boxes of "Single Events" that shall be ignored.
•
Transfer new filter setting(s) to the PROFIBUS device.
Enter menu:
→
Offline menu
Offline menu
Offline menu
Offline menu: "<TAG>
→
Offline
→
Setup
→
Device
→
Diagnosis mapping and
filter
→
Diagnosis filter":
•
All "Single Events" will be displayed which might be filtered.
•
Unmark all check boxes of "Single Events" that shall be ignored.
•
Transfer new filter setting(s) to the PROFIBUS device.
CAUTION!
"Single events" of variable "Event groups" can only be disabled. It is strongly recommended to
use the default setting. If filtering is used please be aware of the fact that the "Single event"
concerned (deactivated by filtering) neither will be taken into consideration during calculation of
both the device status and the status of measurement values nor will be displayed as
information.
Filtering of "Single events" will only be possible if using the VFC 200 PROFIBUS DTM (in
conjunction with PACTware) or the VFC 200 PROFIBUS EDD (in conjunction with SIMATIC PDM).
Filtering of "Single events" will not be possible if using the flowmeter display menu.