5
OPERATION
34
OPTITEMP TT 53 C/R
www.krohne.com
06/2018 - 4006523501 - MA OPTITEMP TT 53 R01 en
Messages in AMS, PDM and FC 475 device diagnostic Actions
Device specific
Failure Error message
Error correction
Device Specific Status (Byte 0-5)
Device Specific Status (Byte 0-5)
Device Specific Status (Byte 0-5)
Device Specific Status (Byte 0-5) Command 48 returns in byte 0 to 13 additional device status information, which are
described below.
Transmitter
M
Underrange
Measurement is under specified range. Check the
specified lower range and sensor and sensor connection.
If error remains replace the sensor.
Transmitter
M
Overrange
Measurement is over specified range. Check the specified
high range and sensor and sensor connection. If error
remains replace the sensor.
Transmitter
F
Ad_short
See Sensor short circuit (same error cause).
Transmitter
F
Ad_open
See Sensor break (same error cause).
Transmitter
F
Sensor break
Sensor wire broken. Check the sensor and sensor
connection. If error remains replace the sensor.
Transmitter
F
Sensor short circuit
Sensor wire shorted. Check the sensor and sensor. If error
remains replace the sensor.
Transmitter
F
Sensor invalid
Invalid sensor selected. Check the sensor selection. If
error remains replace device.
Transmitter
F
Linearization table invalid
Linearization table invalid. Check Linearization table
settings. If error remains replace device.
Transmitter
F
Loop voltage too low
N/A
Transmitter
F
Dac out of range
Digital-to-Analog-Converter out of max range. Power
cycle the device. If error remains replace device.
Transmitter
F
Dac feedback error (+/-
10%)
N/A
Transmitter
F
Loop resistance out of
range
Loop Resistance out of range (RLoad [
Ω
] = (U-8,5)/0.022)).
Check loop connection and voltage. If error remains
replace device.
Transmitter
F
Dac acknowledge error
No acknowledge from Digital-to-Analog-Converter. Power
cycle the device. If error remains replace device.
Transmitter
F
System non-recoverable
errors
Internal system non-recoverable error. Power cycle the
device. If error remains replace device.
Transmitter
M
Process
Maintenance required: Process
Transmitter
M
Configuration
Maintenance required: Configuration
Transmitter
M
Electronics
Maintenance required: Electronics
Transmitter
M
Sensor
Maintenance required: Sensor
Transmitter
S
Process
Out of Specification: Process
Transmitter
S
Configuration
Out of Specification: Configuration
Transmitter
S
Electronics
Out of Specification: Electronics
Transmitter
S
Sensor
Out of Specification: Sensor
Transmitter
C
Process
Check: Process
Transmitter
C
Configuration
Check: Configuration
Transmitter
C
Electronics
Check: Electronics
Transmitter
C
Sensor
Check: Sensor
Transmitter
F
Process
Failure: Process
Transmitter
F
Configuration
Failure: Configuration
Transmitter
F
Electronics
Failure: Electronics
Transmitter
F
Sensor
Failure: Sensor