4
OPERATION
22
MFC 400
www.krohne.com
12/2017 - 4006097501 - AD MFC 400 Profinet R01 en
4.9.6 Unit
4.9.7 Totalizer 1
ID
Name
Data type
Remarks
50
Volume Flow
Unsigned16 Unit selector for volume flow values.
51
Mass Flow
Unsigned16 Unit selector for mass flow values.
52
Flow Speed
Unsigned16 Unit selector for flow speed values.
53
Density
Unsigned16 Unit selector for density values.
54
Temperature
Unsigned16 Unit selector for temperature values.
55
Totalizer 1 - Volume
Unsigned16 Unit selector for totalizer 1 volume values if measurement
(ID 59) is set to "Volume flow".
56
Totalizer 2 - Volume
Unsigned16 Unit selector for totalizer 2 volume values if measurement
(ID 66) is set to "Volume flow".
57
Totalizer 1 - Mass
Unsigned16 Unit selector for totalizer 1 mass values if measurement
(ID 59) is set to "Mass flow".
58
Totalizer 2 - Mass
Unsigned16 Unit selector for totalizer 2 mass values if measurement
(ID 66) is set to "Mass flow".
Table 4-19: Parameter group - Unit
ID
Name
Data type
Remarks
59
Measurement
Unsigned8
Select measurement connected to the totalizer: Volume Flow,
Mass Flow, Concentration Mass Flow 1, Concentration Volume
Flow 1, Concentration Mass Flow 2, Concentration Volume
Flow 2
60
Function
Unsigned8
Off, Absolute Total, Incremental Total, Decremental Total
61
Damping [s]
Float32
Damping for measurement connected to the totalizer, based
on t63.
62
LFC Volume - Threshold
Float32
Low flow cutoff threshold if measurement is set to "Volume
flow".
Unit: selected by ID 50
63
LFC Volume - Hysteresis
Float32
Low flow cutoff hysteresis if measurement is set to "Volume
flow".
Unit: selected by ID 50
64
LFC Mass - Threshold
Float32
Low flow cutoff threshold if measurement is set to "Mass
flow".
Unit: selected by ID 51
65
LFC Mass - Hysteresis
Float32
Low flow cutoff hysteresis if measurement is set to "Mass
flow".
Unit: selected by ID 51
Table 4-20: Parameter group - Totalizer 1