Operation & Installation Manual V12-V6 2010-05-31.doc
31/05/2010
33/60
5.8 Available display information
The information on the display is fully programmable. There are 5 pages available for
displaying different type of information:
•
Two pages with 3 lines for measuring data (page M1 & M2),
•
One page with graphic data (G1),
•
One page with status information (S1)
•
One test page (T1)
By default the display is programmed to show the following information pages:
•
Page M1:
Line 1:
Forward Total
Line 2:
Reverse Total
Line 3:
………………..
•
Page M2:
Line 1:
Process Flow
Line 2:
Process Velocity
Line 3:
Speed of Sound
•
Page G1
Graphics of Process flow over time
•
Page S1
Status/alarm messages
•
Page T1
Test page, the display will cycle through the following screens:
•
KROHNE logo normal
•
KROHNE logo inverted
•
Software version
5.9 Operating the display
In the normal operating mode an information page can be selected by means of the optical
sensitive buttons “up” (
▲
) and “down” (
▼
) (scrolling mode).
The information pages (lay out and content) can be changed in the programming mode. To
enter the programming: touch and hold the button > as long as the display tells to hold this
button and release this button when the display tells to release it.
In programming mode a menu structure is entered, the display will show a menu number (in
the upper right corner) and the display will show three or four lines
Three lines (applicable when the current item in the option list is a submenu):
•
Preceding item in options list
•
Current item in options list
•
Next item in options list
Four lines (applicable when the current item in the option list is a menu item that can be
changed or assigned a value to):
•
Preceding menu item
•
Current menu item
•
Setting/value of current menu item
•
Next menu item