31
KPS-PA60 • Digital clamp meter
ENG
•
Operating altitude: max. 2000 meters (7000 ft.)
•
Display: 4000 counts with analog bar LCD display
•
Maximum value display: 4000 digits
•
Polarity indication: automatic; „-„ for negative polarity.
• Overrange indication: „0L‟ or „-0L‟
•
Converter Rate: 3 times/sec; Bar graph: 30 times/sec.
•
Unit indication: function and unit.
•
Auto power off time: 15 minute.
• Operating power : 1.5V×3 AAA batteries
•
Battery low indication:‘
’ on LCD
• Temperature factor: < 0.1×Accuracy / ºC
• Operating temperature: 0ºC to 40ºC (32ºF to 104ºF)
• Storage temperature: -10ºC to 50ºC (10ºF to 122ºF)
• Dimension: 225×86×33mm
•
Weight: approximate 320g (including batteries)
3.2 Electrical specifications
Temperatura ambiente: 25±3°C, humedad relativa (HR): <75%
3.2.1 AC Current
Range
Resolution
Accuracy
40A
0,01 A
±(2,5% of r 6 digits)
400A
0,1 A
1000A
1 A
- Max. input current: 1000A AC
- Frequency range: 40 to 100Hz
3.2.2 DC Current
Range
Resolution
Accuracy
40A
0,01 A
±(3,0% of r 6 digits)
400A
0,1 A
1000A
1 A