34
KPS-PA60 • Digital clamp meter
ENG
3.2.5.3 By Hz/DUTY range:
Range
Resolution
Accuracy
9,999Hz
0,001Hz
±(0,5% of r 5 digits)
99,99Hz
0,01Hz
999,9Hz
0,1Hz
9,999kHz
0,001kHz
99,99kHz
0,01kHz
999,9kHz
0,1kHz
9,999MHz
0,001MHz
- Overload protection: 250V dc or 250V ac rms.
- Input Voltage range: 200mV-10V ac rms
3.2.6 Duty Cycle
Range
Resolution
Accuracy
0,1 – 99,9%
0,1%
±3,0%
•
By A range ( from current clamp):
- Frequency response: 10 ~1kHz
- Input current range: ≥70A rms AC
- Max. input current: 1000A
•
By ACV range:
- Frequency response: 10 ~ 10 kHz
- Input voltage range: ≥ 1 V rms AC
- Input impedance: 10MΩ
- Max. input voltage: 600V rms AC
• By Hz/DUTY range
- Frequency response: 1 ~ 10MHz
- Input voltage range: ≥ 500mV rms
- Overload protection: 250V rms AC