974-72007001/1.0 1007D
Altimeter Operators Manual
Page
26 Telemetry
Kongsberg Mesotech Ltd.
Port Coquitlam, BC - Canada
'G'm
–
Set Auto-Gain-Adjust Mode
This command selects the TVG adjustment method
used if auto-gain is enabled by the
Set Threshold
command.
m
is one of:
'0'
Disabled (regardless of the
Set Threshold
command setting)
'1'
Add or subtract a flat offset to shift the
TVG curve up or down.
The performance and stability of the Auto-Gain-
Adjustment is strongly dictated by:
Settings that affect the ping rate (e.g.
Set Range
command)
Bottom Detection Method used (
Set Detection
Method
command)
Deployment conditions (acoustic noise, multipath,
turbulence, water column sediment, etc).