![Kobold KEC-1 Operating Instructions Manual Download Page 13](http://html1.mh-extra.com/html/kobold/kec-1/kec-1_operating-instructions-manual_1982459013.webp)
Measuring range
KEC-1
K01/1216
Page
13
of
43
4.2.2 Measuring range „Max Version“
Innerdiameter of pipe
Max Version (185,0 m/s)
Full scale value in in Nm³/h
Recommanded
Sensor lenght
Inch mm
Air
2)
Air
3)
Ar
3)
CO2
3)
N2
3)
O2
3)
N2O
3)
Methane
3)
Natural
gas
3)
1/2"
16.1
90.98 83.7 142.2 90.0 80.9 86.7 89.2 53.8
160 mm
3/4"
21.7
177.84 163.5 278.0 176.0 158.1 169.6 174.4 105.2
1" 25.0
243.88 224.2 381.2 241.4 216.9 232.5 239.1 144.2
26.0
265.20 243.8 414.6 262.5 235.8 252.9 260.0 156.8
27.3
294.72 271.0 460.7 291.7 262.1 281.0 289.0 174.3
28.5
323.32 297.3 505.4 320.0 287.5 308.3 317.0 191.2
30.0
361.08 332.0 564.5 357.4 321.1 344.3 354.1 213.5
1 1/4"
32.8
436.69 401.5 682.7 432.2 388.3 416.4 428.2 258.2
36.0
531.48 488.7 830.8 526.0 472.6 506.8 521.1 314.3
36.3
541.06 497.5 845.8 535.5 481.1 515.9 530.5 320.0
1 1/2"
39.3
639.84 588.3 1000.2 633.3 568.9 610.1 627.4 378.4
40.0
663.68 610.2 1037.5 656.9 590.1 632.8 650.8 392.5
41.9
728.41 669.7 1138.7 720.9 647.7 694.5 714.2 430.7
43.1
777.34 714.7 1215.2 769.4 691.2 741.2 762.2 459.7
45.8
882.17 811.1 1379.0 873.1 784.4 841.2 865.0 521.7
2"
50.0
1059.23 973.9 1655.8 1048.3 941.9 1010.0 1038.6 626.4
51.2
1112.05 1022.5 1738.4 1100.6 988.8 1060.4 1090.4 657.6
53.1
1197.59
1101.1
1872.1
1185.3
1064.9
1141.9
1174.3
708.2
54.5
1263.13 1161.4 1974.6 1250.2 1123.2 1204.4 1238.5 747.0
57.5
1489.43 1369.5 2328.3 1474.1 1324.4 1420.2 1460.5 880.8
60.0
1544.12 1419.8 2413.8 1528.3 1373.0 1472.3 1514.1 913.1
64.2
1774.33 1631.4 2773.7 1756.1 1577.7 1691.8 1739.8 1049.3
2)
Referred to DIN 1945 / ISO 1217 (20°C, 1000mbar) and compressed air.
3)
Referred to DIN 1343: 0°C, 1013.25 mbar