Installation
Next steps
Finish the installation and perform the initial startup.
Analog Control
Analog ports serve for exchanging analog control signals. Refer-
ence point for the signals is the connector GND.
OUT: Device sends signal.
IN: Device receives signal.
Power Supply
Note
The nominal capacity of the connected devices must be maxi-
mum 50 % of the power supply to account for larger inrush cur-
rents when switching on the modules.
The maximum power input is 100 watts.
Prerequisites
The electrical power supply at the installation site must be
connected directly to the nearest main power line.
The power must be free from ripple, residual current, voltage
peaks and electromagnetic interference.
The connectors for the mains voltage are grounded accord-
ingly.
The device receives sufficient power with reserve capacity.
The device is intended for use with AC power networks of 100–
240 V.
Power cable
Use only the enclosed power cable to connect the device to the
mains. Replace defective power cables only with accessories
from KNAUER. Only use power cables admitted for use in your
country.
Procedure
Process
Figure
1. Place the terminal strip
3
on a suitable surface.
2. Push the depressor tool
1
into the opening on the
upper side.
3. Continue pushing the
depressor tool down and
lead the cable
2
into the
front end of the terminal
strip.
4. Remove the depressor
tool.
5. Check whether the cables
are tightly attached.
6. Plug the terminal strip
onto the multi-pin con-
nector.
Fig. 29
terminal strip
1
2
3