KNAUER
P 6.1L User Manual V6890 Version 1.2
33
Installation
Wiring the Terminal Strip
To control one device through another, you use the multi-pin
connector. To use remote control, you have to connect cables to
the terminal strip (everything comes included with delivery).
The single ports are used to exchange control signals.
Prerequisites
The device has been turned off.
The power plug has been pulled.
Tools
Depressor tool
CAUTION!
Short-circuit hazard.
- Turn off the device before connecting it to the
multi-pin connector.
- Pull the power plug.
CAUTION!
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic
discharge and ground.
Relay contact
The contact is on a floating basis. Its
setting depends on the settings in the
Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least
1000 ms
Permissible load of the relay contact:
1 A/ 24 V DC
Relay contact
The contact is on a floating basis. Its
setting depends on the settings in the
Control Unit or software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least
1000 ms
Permissible load of the relay contact:
1 A/ 24 V DC
Assignment
Connection
Function