KNAUER
P 6.1L User Manual V6890 Version 1.2
25
Installation
Connecting the Leak Management
The leak management consists of the leak sensor and the drain-
age system. The drainage system ensures that escaping liquids
flow into a waste bottle. If there is too much liquid, the red LED
starts flashing. Both the device and the data acquisition via chro-
matography software are stopped.
Prerequisite
The front cover has been removed.
Procedure
Process
Figure
1. Plug one silicone tube
onto the inlet
3
of the
flush pump and con-
nect to the flushing
solution bottle.
2. Plug a second silicone
tube onto a vacant cap-
illary connector
1
of
the flush pump and
connect to the flushing
solution bottle.
Fig. 19
Silicone tube
between
piston backflushing and
flush pump
1
2
3
Procedure
Process
Figure
1. Carefully push the fun-
nel
1
into the center
opening of the capillary
guide
2
.
Abb. 20 Funnel and capillary
guide
2. Push the long ending of
the first nozzle
4
into
the hose
3
.
Abb. 21 Hose and nozzle
1
2
3
4