SMY133-SMY134-SMP133 Operating Manual
So, when you want to change anything from the instrument panel, you must :
1. choose the user (
Peter
in our example)
2. enter the PIN corresponding this user in the user management setup
For details see the application note No. 004 :
Users, passwords and PINs
.
In case the PIN is lost, ask manufacturer for alternate PIN via manufacturer's website at
4.2.13 Instrument Information
•
Instrument Model
&
Serial Number
… Instrument hardware model & serial No.
•
Instrument Hardware, Firmware
&
Bootloader Versions
…Instrument hardware &
firmware specification.
•
Object Number
… Measured node specification ( preset by ENVIS-DAQ program for data
identification ).
•
Vbatt
… Backup battery voltage (if equipped)
•
Error Code
… Indicates some instrument's hardware or setting problem. At normal state
equals to 0. In case of detection of any error it contains a number created as the sum of
binary weights of the errors. The following table provides overview of them and
recommended action:
Tab 4.1 : Instrument Errors
Error No.
(weight)
Error
Action
1
RAM memory error
set the instrument (optimally with the ENVIS program, if
possible) to the
default setting
; if the error appears again
send the instrument to a service organization for repair
2
instrument setup error set the instrument (optimally with the ENVIS program, if
possible) to the
default setting
; if the error appears again
send the instrument to a service organization for repair
4
calibration error
the instrument must be recalibrated – send to a service
organization
8
wireless
communication module
error (Wifi/Zigbee)
send to a service organization
16
RTC error
in
the time setup window
(or better with the ENVIS program, if
possible), set actual date&time; if the error appears again,
check built-in backup battery; otherwise send the instrument
to a service organization for repair
128
archive data error
clear all of the archives with the ENVIS program; if the error
appears again send the instrument to a service organization
for repair
256
FLASH memory error
send to a service organization