Electrosurgical Unit Minicutter
Revision 08
39
Contact coagulation
Coagulation with high depth upon direct contact between electrode and tissue
Power setting range:
Max. 70 W at 200
Ω
Max. periodic HF output voltage:
Max. 1,300 V
max
Max. output current
600 mA at 200
Ω
Crest factor
5.0
Danger of injury from excessive voltage!
Active accessories to be used with this current must be able to withstand an HF voltage of at least 1,300 V.
It must be kept in mind that voltage is achieved even at settings below 10, and that the active accessory
must be selected accordingly.
Performance Diagram
H
F
po
we
r i
n
W