6.9
Special Scanner Functions
6.9.1
Continuous Scan
•
CYCLE PAUSE/CYCLE RATE
Setting: OFF
When setting a cycle pause of less than 10 ms or a cycle rate of more than 5 Hz, OFF appears
on the display and continuous scanning without pause is activated as long as the foot switch is
pressed.
6.9.2
CYCLE
•
CYCLE PAUSE (Setting
TIME MODE
)
Setting range: 10 ms–10 s
•
CYCLE RATE (Setting
FREQUENCY MODE
)
Setting range: 0.1–5 Hz
Fig. 6-10:
Setting CYCLE SCAN, here in Frequency Mode
When selecting the repetition function
CYCLE
for multiple use of the scan shapes, the time
between 2 scans (Time Mode) or the repetition frequency (number of scans per second,
Frequency Mode) appears in the menu line D5. The scan shape is repeated as long as the foot
switch is activated.